GENERAL REQUIREMENTS
- Type of therapeutic radiation: electrons
- Number of electron therapeutic energies (min. 3)
- The lowest energy of the therapeutic beam max. 6 MeV
- The highest energy of the therapeutic beam min. 12 MeV
- Built-in motorized beam stopper, mechanically interlocked to the head position
- The beam stopper should automatically follow the movements of the head with an accuracy of 1 cm, including when underneath the surgical table or obscured by surgical drapes
- Built-in, autonomous cooling system
Internal shielding limits stray radiation to 0.3 uSv/Gy at 3m distance from the patient plane - Device Mobility – Built in moving and steering system to easily move a complete unit between operating room
- Maximum weight of system running (portable) parts including beam stopper: max. 1 500 kg
- Soft Docking (no contact between the applicator and treatment head)) alignment system.
- The docking system shall use cameras or lasers, shall be interlocked to assure proper alignment and shall automatically assist in the alignment process.
- System shall comply and be consistent with AAPM Task Group 72 for mobile IORT electron linacs
- At least ten (10) operational linacs of the same model and energy range in clinical operation around the world
International Publication for Breast:
International Publication for Colorectal:
International Publication for Superficial:
Flash Therapy:
- European consensus BCC_2019

- NCCN Guidelines – 2020 – MCC

- NCCN Guidelines – nmsc 2020 – Squamous

- Skin RT consensus guidelines ASTRO Jan 2020

- Treatments of Primary BCC of the Skin – Review and Meta-analysis_2018

- Radiation Research Special issue on FLASH_DECEMBER 2020

- Schuler – Experimental Platform for Flash using Clinical Linac

- 2000 Biochemistry 29 10978-89

- FLSAH RT and oxygen depletion_Petersson_2020

- HOSEY_et_al-1966-Nature